valve or DO to multiple trains, but you can configure the same detector to multiple
trains.
4. On the
Application
→
Timed Events
screen, filter the configured events as per train
selection by selecting the
Train#
check box.
4.4.17
Use this feature to configure a single analysis clock or multiple clocks.
One analysis clock can be considered as one virtual gas chromatograph (GC) that has
independent
Sample Loop
,
Analytical Path
, and
Timed Event
tables.
Multiple analysis clocks can run independently to analyze multiple streams at the same
time. Emerson sets the number of analysis clocks at the factory per the mechanical
configurations of the GC.
Mechanical
configurations
Description
Trains (1 - 6)
The configured trains that are used by the analysis
Default Stream
Sequence (Def
Strm Seq)
Sets the default sequence to be used by the indicated analysis during auto-
sequencing.
Purge Duration
The amount of time, in seconds, to purge the stream before starting an
analysis, calibration, or validation run. The default value is
60 SEC
. Purging
allows sample gas to flow through the sample loop prior to beginning the
run.
Energy Value
Check
If enabled, the GC analyzes the calibration gas as an unknown stream and
computes its energy value. The GC then compares this value to the
Cal Gas
Cert CV
and determines if the calibration gas's energy value is within the CV
Check Allowed Deviation. If it isn't, the GC triggers the
Energy Value Invalid
alarm. The following conditions must be met before the GC can perform a EV
Check:
•
The
EV Check
flag on the
System
screen must be enabled.
•
At least one stream must be set up in the
Streams
screen as a calibration
stream, and the
Auto
flag for this stream must be enabled.
The EV Check is performed under any of the following circumstances:
•
During a warm start that follows a power failure during normal operation.
The GC waits for the heater and electronic pressure controller (EPC) to
reach their respective set points and stabilize. It then analyzes the
calibration gas as an unknown stream and identifies the peaks. If all the
component peaks are identified, the GC computes the calibration gas'
energy value and performs the EV Check.
•
After a successful calibration, the GC computes the gas's energy value
with the new response factors and performs the EV Check
1. Press
Insert
to add a new analysis.
2. Press
Delete
to delete an analysis.
Installation and start-up
Reference Manual
June 2022
2-3-9000-744
94
Emerson.com/Rosemount
Summary of Contents for Rosemount 700XA
Page 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Page 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Page 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Page 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Page 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Page 280: ......
Page 281: ......
Page 282: ......
Page 293: ......