Figure C-2: µFPD PMT
C.1
Configure the micro flame photometric
detector (µFPD)
Procedure
1. Open MON2020 and select
Hardware
→
Detectors
.
If your gas chromatograph (GC) has an integral FPD,
FPD G2
displays in Slot 1 or Slot
2.
Note
Before making any modifications to this window, halt the analysis.
2. Select
Hardware
→
Detectors
.
3. Select
Manual
if you want to control the burner ignition; select
Auto
if you want the
GC to control the burner ignition.
The following data displays for each detector:
Name
Description
Det #
Numerical identifier for the detector to which the following data applies.
Detector
Options, which depend on your GC's configuration, are:
•
TCD (thermal conductivity detector)
•
FPD G2 (µFPD)
•
FID (flame ionization detector)
•
FID G2
H2 Valve
Optional hydrogen carrier shut-off valve
Flame Temp RTD
Select the appropriate resistance temperature device (RTD) from the drop-down
list. The RTD measures the temperature of the flame.
Micro flame photometric detector (µFPD)
Reference Manual
June 2022
2-3-9000-744
268
Emerson.com/Rosemount
Summary of Contents for Rosemount 700XA
Page 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Page 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Page 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Page 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Page 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Page 280: ......
Page 281: ......
Page 282: ......
Page 293: ......