Procedure
1. Start MON2020 and connect to the GC.
2. Select
Applications
→
Communication
.
The
Communication
window displays.
3. Select the appropriate mode from the
Port
drop-down list for the appropriate serial
port.
The options are:
•
RS232
•
RS485
•
RS422
Figure 5-50: Rosemount MON2020 Communication window
4. Click
OK
.
5. Close MON2020.
6. Turn off the GC.
7. Locate and remove the base in/out (I/O) board, which is located in the card cage in
the GC’s lower enclosure.
8. Consult
, which shows the correct switch settings.
Operation and maintenance
Reference Manual
June 2022
2-3-9000-744
166
Emerson.com/Rosemount