The Rosemount 700XA has four serial communication ports: Port 0, Port 1, Port 2, and
Port 3, which is a dedicated personal computer (PC) to gas chromatograph (GC) port.
You can set the mode for each of the first three ports to RS232, RS422, or RS485.
Normally, the customer specifies these port configurations at the time of order, and
Emerson sets them at the factory. However, the customer can use Rosemount MON2020
to change them at any time.
Note
The backplane has two switches located at
SW1
. The first switch is used for starting the
DHCP server. The second switch is reserved for future use.
The backplane has two Ethernet ports:
Name
Location
Connector type
ETHERNET1
J22
RJ45 (DHCP-enabled)
ETHERNET2
TB11
4-wire terminal block
Related information
Connect directly to a personal computer (PC) using the gas chromatograph's (GC’s)
Ethernet1 port
Maximum distance by communication type
Communication Type
Maximum distance
RS-232
50 ft (15.24 m)
RS-422/RS-485
4,000 ft (1,219.20 m)
Ethernet (CAT5)
300 ft (91.44 m)
Operation and maintenance
Reference Manual
June 2022
2-3-9000-744
164
Emerson.com/Rosemount
Summary of Contents for Rosemount 700XA
Page 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Page 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Page 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Page 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Page 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Page 280: ......
Page 281: ......
Page 282: ......
Page 293: ......