WARNING
Explosion
Failure to de-energize the analyzer may cause an explosion and severely injure personnel.
Before opening the analyzer, disconnect all electrical power and ensure that the area is
free of explosive gases.
Keep cover tight while circuits are live.
Use cables or wires suitable for the marked "T" ratings.
Cover joints must be cleaned before replacing the cover.
Conduit runs to the enclosure must have sealing fitting adjacent to enclosure.
Procedure
1. Disconnect all power to the GC.
2. If you have not already done so, remove the explosion-proof dome and the thermal
cover.
Reference Manual
Operation and maintenance
2-3-9000-744
June 2022
Rosemount 700XA
121