The GC-to-PC interface provides you with the greatest capability, ease-of-use, and
flexibility. You can use Rosemount MON2020 to edit applications, monitor operations,
calibrate streams, and display analysis chromatograms and reports, which can then be
stored as files on the PC’s hard drive or printed from a printer connected to the PC.
WARNING
Hazardous area explosion hazard
Failure to follow this warning may result in injury or death to personnel.
Do not use a personal computer (PC) or printer in a hazardous area.
Emerson provides serial and Ethernet communication links to connect the analyzer to
the PC and to connect to other computers and printers in a safe area.
1.1.3
Sample conditioning system (SCS)
The optional sample conditioning system is located between the process stream and the
sample inlet, which is often mounted below the gas chromatograph (GC).
The standard SCS configuration includes a stream switching system and filters.
1.2
Functional description
A sample probe installed in the process line takes a sample of the gas to be analyzed from
the process stream. The sample passes through a sample line to the sample conditioning
Overview
Reference Manual
June 2022
2-3-9000-744
10
Emerson.com/Rosemount
Summary of Contents for Rosemount 700XA
Page 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Page 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Page 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Page 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Page 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Page 280: ......
Page 281: ......
Page 282: ......
Page 293: ......