The input range for an AC power supply is 90 to 264 volts (auto-ranging).
6.10
Voltage LEDs
A set of LEDs can be found above the test points .
These LEDs are a quick way to visually inspect the voltage status of some of the gas
chromatograph's (GC's) electrical components.
Figure 6-2: Voltage LEDs
The following LEDs are associated with the following GC components:
VIN (fuse open)
Glows red when the fuse has blown or been removed;
otherwise, it is not lit.
1 RTN - 24 loop
(Power)
Glows when the power supply for the analog outputs is
functioning properly; otherwise it is not lit.
24V regulated (GC
power)
Glows when the GC power is functioning properly;
otherwise, it is not lit.
17V (Input for the
preamp)
Glows when the power supply for the preamp is
functioning properly; otherwise, it is not lit.
12V (Input for the
I/O cards0
Glows when thee optional ROC expansion card's power
supply is functioning properly; otherwise, it is not lit.
5V1
Glows when the system chip's 5.1 V power supply is
functioning properly; otherwise, it is not lit.
3V4
Glows when the 3.4 V power supply for the system chips
is functioning properly; otherwise, it is not lit.
Power ON
Glows when the GC is on; otherwise, it is not lit.
Reference Manual
Troubleshooting
2-3-9000-744
June 2022
Rosemount 700XA
195
Summary of Contents for Rosemount 700XA
Page 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Page 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Page 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Page 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Page 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Page 280: ......
Page 281: ......
Page 282: ......
Page 293: ......