8. Disconnect the wire terminals for J6, J10, and J2.
Figure 5-34: µFPD PMT module connections
A. J6 - CAN and power signals
B. J10 - igniter
C. J2 - thermocouple
D. USB connector
E. PMT module bracket screws
F. Stream manifold screws
G. Stream manifold
9. Loosen the two Phillips screws from the stream solenoid manifold.
10. Move the entire manifold away from the µFPD PMT module.
11. Use a large flat head screwdriver to loosen the two µFPD PMT module bracket
screws.
12. Move the 1/16-in. stainless steel tubing away and pull the µFPD PMT module out of
the enclosure.
Install the micro flame photometric detector (µFPD) photo
multiplier tube (PMT) module
Tools required
•
7 ⁄ 64-in. Allen wrench
•
Phillips screwdriver
•
Flathead screwdriver
Procedure
1. Lower the µFPD PMT module into the enclosure.
Operation and maintenance
Reference Manual
June 2022
2-3-9000-744