5. Check columns.
6.5
Small peaks
Recommended actions
1. Clean the burner windows with isopropyl alcohol (IPA).
2. Adjust the air to H
2
flow rate to between 1.3 and 1.7.
3. Ensure that the flame photometric detector (FPD) high voltage is 800 Vdc for
side cart option of FPD.
6.6
No temperature readings
Recommended action
Check the thermocouple.
The thermocouple's resistance should be less than 3.5 ohms.
6.7
Noisy baseline
Recommended action
Click
Autozero
to adjust the baseline.
6.8
Peak clipping
Recommended action
Click
Autozero
to adjust the baseline.