4 Repair
4.3 Replacing parts (without removing covers)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
98
66 81 832 D3696
D3696.076.01.02.02 05.2019
9.
Screw on the partition plate (2 screws).
10. If the option is available, install the battery again (see "Replace
11. Close the service cover (see "Closing the service cover [→ 52]").
4.3.2.4.1
Replace PC power supply
ü
The PC is removed (see "Replacing the PC including the power
supply unit [→ 94]").
1.
Turn the PC on its side.
2.
Remove the 2 fastening screws.
3.
Pull out the power supply unit out of the guide tabs. In doing so,
make sure the foam rubber strip (A) does not become damaged.
4.
Disconnect the 4 plug connections and dispose of the defective
power supply unit.
5.
Connect the 4 plug connections from the new power supply unit
together.