4 Repair
4.3 Replacing parts (without removing covers)
66 81 832 D3696
D3696.076.01.02.02 05.2019
89
8.
Re-tighten the locking plate mini DIN connector with 1 screw (A).
9.
Route the cables as shown and clip them into the holders.
10. Close the service cover (see "Closing the service cover [→ 52]").
4.3.2.3
Control unit
4.3.2.3.1
Removing and inserting the control unit
Removal
ü
The monitor is removed; see "Replacing the monitor [→ 85]".
1.
Unscrew the fastening screws (A) from the locking plates.