4 Repair
4.2 Removing covers and holders
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
58
66 81 832 D3696
D3696.076.01.02.02 05.2019
4.2.2
Replacing the CEREC Primescan / Primescan scanner
holder
A
CEREC Primescan /
Primescan scanner holder
cover parts set
B
CEREC Primescan / Primes-
can scanner holder
C
CEREC Primescan /
Primescan internal scan-
ner cable
ü
The service cover is open (see "Opening the service cover [→ 52]").
ü
The right cover is removed (see "Right cover [→ 54]").
1.
Detach the scanner connector from the scanner cable and set the
scanner aside safely.
2.
Unscrew the 3 screws (A) from the device and remove the scanner
holder laterally from the device.
3.
Insert the new scanner holder and firmly screw the three screws
onto the device .