![Dentsply Sirona CEREC Omnicam AC Service Manual Download Page 104](http://html.mh-extra.com/html/dentsply-sirona/cerec-omnicam-ac/cerec-omnicam-ac_service-manual_2483632104.webp)
4 Repair
4.3 Replacing parts (without removing covers)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
104
66 81 832 D3696
D3696.076.01.02.02 05.2019
4.3.2.10
Replacing the WLAN card
1.
Open the PC (see "Opening the PC [→ 102]").
2.
Remove the screw from the PCI slot plate and pull the WLAN card
up and out of the PC.
3.
Insert the new WLAN card in the slot.
4.
Fasten the slot plate with a screw.
5.
Close the PC (see "Closing the PC [→ 107]").
Note on connecting the WLAN antennae
The WLAN card may have two or three antenna ports. If three antennae
are installed in the acquisition unit but the WLAN card only has two
ports, connect one antenna from each side of the acquisition unit. For
this use the antenna closest to the PC ports on the side that has two
antennae. Next secure the antenna cable for the third antenna not in
use onto the enclosure using a cable tie so that it is not hanging freely.
4.3.2.11
Replacing the interface card
1.
Open the PC (see "Opening the PC [→ 102]").
2.
Remove the screw from the PCI slot plate and pull the interface
card up and out of the PC.