![Dentsply Sirona CEREC Omnicam AC Service Manual Download Page 125](http://html.mh-extra.com/html/dentsply-sirona/cerec-omnicam-ac/cerec-omnicam-ac_service-manual_2483632125.webp)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
5 Settings and adjustments
5.1 Set the monitor kinematics
66 81 832 D3696
D3696.076.01.02.02 05.2019
125
9.
Carefully pivot the monitor holder (C) in order to get past the
kinematics bearing flange with the lever arm cover at the front (D).
NOTE
Cables may become crushed
Ensure that the lever arm covers fit into each other perfectly and that
no cables are caught in between.
10. Pivot the entire monitor holder away from you and screw the 4 PT
screws (B, F 3 X 8) in finger tight. NOTE! Do not tighten too hard
as otherwise the screw bosses may burst!