![Dentsply Sirona CEREC Omnicam AC Service Manual Download Page 79](http://html.mh-extra.com/html/dentsply-sirona/cerec-omnicam-ac/cerec-omnicam-ac_service-manual_2483632079.webp)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
4 Repair
4.3 Replacing parts (without removing covers)
66 81 832 D3696
D3696.076.01.02.02 05.2019
79
2.
Unscrew the 2 M12 nuts (A) and screw them away from the bearing
pedestal.
3.
Apply Loctite carefully on the gear shaft between the M12 nuts and
bearing pedestals.
4.
Screw the 2 M12 nuts back onto the bearing pedestal.
5.
Loosen the 2 screws (B, M4 X 35). Leave the other ones screwed in
so that the centering does not change!
6.
Retighten one M12 nut (A) in a clockwise direction (move wrench
up). Start with a 1/2 rotation.
Ä
The disk springs are pressed together -> the tightness
increases.