4 Repair
4.3 Replacing parts (without removing covers)
66 81 832 D3696
D3696.076.01.02.02 05.2019
69
6.
Carefully disconnect the connector (E) from the camera and remove
the camera cable.
7.
Detach the spacer (F) and the camera cover from the old camera
cable and push it over the new camera cable in the order
presented.
Pay attention to the correct alignment of the spacer, as depicted in
the illustration.
The old camera cable is no longer required and can be disposed of.
8.
Insert the connector (E) of the new camera cable into the socket of
the camera.
9.
Place the hex (G) of the new camera cable with a flat side
downwards onto the EMC adhesive pad (H) of the camera housing.
10. Place the strain relief cover (D) onto the camera housing and fix it
with the two screws (tightening torque: 30 Ncm).