3 Scanner
3.1 Calibrating the scanner
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
36
66 81 832 D3696
D3696.076.01.02.02 05.2019
Measuring the position of the exit window
1.
Mount the bottom side of the calibration set to the tip of the scanner.
2.
Click on the
"OK" button.
Ä
The calibration process is continued.
Ä
Once the calibration is complete, a message is displayed
indicating this.
3.
Confirm the message by clicking the
"OK" button.
Ä
The CEREC Primescan / Primescan / CEREC Omnicam / Omnicam
scanner is calibrated.
Error message during calibration
The software indicates if an error occurs during calibration. If the
calibration process resulted in errors, restart the process.
End calibration
ü
The software indicates that the calibration was completed
successfully.
➢ Click the
"OK" button.
Ä
The CEREC Primescan / Primescan / CEREC Omnicam /
Omnicam scanner is calibrated.