Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
4 Repair
4.2 Removing covers and holders
66 81 832 D3696
D3696.076.01.02.02 05.2019
57
3.
Press the 2 Pilz connectors (C) from behind out of the grommets.
Alternatively: Loosen the cover by pulling on the bottom part of the
front cover until the Pilz connector audibly disengages.
4.
Paying attention to the scanner cable, press the front cover
downwards slightly and pull it forwards and out.
4.2.1.7
Replacing the kinematics cover
1.
Unscrew 2 screws and remove cover bearing flange.
2.
Unscrew 4 screws and remove bottom arm cover.
3.
Unscrew 2 screws from cable fixings and remove.
4.
Remove front arm cover; it may be necessary to move the
kinematics a little to do so.
5.
Unscrew 2 screws and remove monitor bearing cover.
6.
Reassembling everything in reverse order, when doing so note the
following. If possible, fix cable back in the cover in the same
position in which it was found. Exercise great care in this process.
During the swiveling movement, neither may tension be applied to
the cable nor may protruding cable inhibit the swiveling movement.