Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
4 Repair
4.3 Replacing parts (without removing covers)
66 81 832 D3696
D3696.076.01.02.02 05.2019
83
6.
Pivot the cable reliefs above the cables and screw these in with the
2 PT screws (E, F 3 X 8) finger tight. NOTE! Do not tighten too
hard as otherwise the screw bosses may burst!
7.
Carefully retighten the cables. These must run above each other
under the cable guard and may only be pressed slightly by the
cover.
8.
Insert the lever arm cover at the front.
9.
Carefully pivot the monitor holder (C) in order to get past the
kinematics bearing flange with the lever arm cover at the front (D).
NOTE
Cables may become crushed
Ensure that the lever arm covers fit into each other perfectly and that
no cables are caught in between.