![Dentsply Sirona CEREC Omnicam AC Service Manual Download Page 112](http://html.mh-extra.com/html/dentsply-sirona/cerec-omnicam-ac/cerec-omnicam-ac_service-manual_2483632112.webp)
4 Repair
4.4 Replacing parts (whereby the removal of covers is necessary)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
112
66 81 832 D3696
D3696.076.01.02.02 05.2019
4.4.4
Replacing the fan power pack.
ü
The right cover is detached (see "Removing and installing covers
[→ 51]").
ü
The PC is removed (see "Replacing the PC including the power
supply unit [→ 94]").
1.
Pull the 4 fastening pins inwards and out and detach the fan.
2.
Hold the new fan over the fan opening.
3.
Pull the 4 supplied fastening pins into the holes on the chassis and
on the fan until the heads are on the inside of the chassis.
4.
Route the cables through the opening into the chassis interior.
5.
Shorten the 4 fastening pins so that around 10 mm protrude at the
fan.
6.
Reinstall the right cover (see "Removing and installing covers
[→ 51]").
7.
Reinstall the PC (see "Replacing the PC including the power supply
unit [→ 94]").