![Dentsply Sirona CEREC Omnicam AC Service Manual Download Page 106](http://html.mh-extra.com/html/dentsply-sirona/cerec-omnicam-ac/cerec-omnicam-ac_service-manual_2483632106.webp)
4 Repair
4.3 Replacing parts (without removing covers)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
106
66 81 832 D3696
D3696.076.01.02.02 05.2019
6.
Press the socket lock downwards.
7.
Pull the graphics card up and out of the PC.
8.
Insert the new graphics card into the slot and lock it.
9.
Fasten the slot plate with 2 screws.
10. Insert the connector until it latches.
11. Screw the attachment bracket back on with 2 screws (A).
12. Close the PC (see "Closing the PC [→ 107]").