4 Repair
4.3 Replacing parts (without removing covers)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
86
66 81 832 D3696
D3696.076.01.02.02 05.2019
2.
Loosen and remove the monitor brake (2 screws).
3.
Unscrew the locking plate mini DIN connector (1 screw A).
4.
Detach the three cables that run from the monitor to the PC from
the PC and unscrew the ground cable.
5.
Release the monitor cables from the clips.
6.
Remove the monitor in an upward motion.