![Dentsply Sirona CEREC Omnicam AC Service Manual Download Page 114](http://html.mh-extra.com/html/dentsply-sirona/cerec-omnicam-ac/cerec-omnicam-ac_service-manual_2483632114.webp)
4 Repair
4.4 Replacing parts (whereby the removal of covers is necessary)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
114
66 81 832 D3696
D3696.076.01.02.02 05.2019
4.4.6
Replacing the footswitch cable
ü
The entire cover is detached (see "Removing and installing covers
[→ 51]").
1.
Unscrew the switch (2 screws).
2.
Unplug the connector at the other end of the cable and take the
cable out.
3.
Insert the new footswitch cable and attach the connector.
4.
Screw the switch back on with 2 screws.
5.
Adjust the footswitch's switching point with the adjusting screw (see
"Adjusting the footswitch [→ 62]").
6.
Reinstall the covers (see "Removing and installing covers [→ 51]").
4.4.7
Replace the handle
ü
The control unit is removed (see "Removing and inserting the
control unit [→ 89]").
1.
Remove the four screws and then take off the handle.
2.
Take the new handle and push the 8 spacers into the holes in the
handle from both sides.
3.
Screw on the new handle with the four screws.
4.
Reinstall the control unit (see "Removing and inserting the control
unit [→ 89]").