2 Service tools
2.2 "OTC" radio module tool
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
22
66 81 832 D3696
D3696.076.01.02.02 05.2019
2.2
"OTC" radio module tool
Background
If the pairing between two radio modules is defective, the OTC tool can
be used to reset the acquisition unit radio module via software. Please
note that a controlled pairing requires that the radio module to be paired
is also reset.
Reset radio module
ü
The LAN cable for the radio module is plugged in.
1.
Start the "OTC.bat" tool. The program is preinstalled on the
acquisition unit and can be found at C:\DentsplySirona\Tools\BM-
Radio.
2.
Connect with the installed radio module by activating the
"Connect"
button.
3.
Press the
"Reset"
button to reset the radio module.
Ä
The radio module is reset.