![Dentsply Sirona CEREC Omnicam AC Service Manual Download Page 119](http://html.mh-extra.com/html/dentsply-sirona/cerec-omnicam-ac/cerec-omnicam-ac_service-manual_2483632119.webp)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
5 Settings and adjustments
5.1 Set the monitor kinematics
66 81 832 D3696
D3696.076.01.02.02 05.2019
119
10. Unscrew 2 screws (F, M3 X 6 Precote) and remove the 2 monitor
connection covers (G).
5.1.2
Adjust the monitor kinematics
1.
Guide the 4 monitor cables through the sleeve bearing and insert
the monitor into the acquisition unit without any cover on the
monitor holder.