Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
8 Re-test
8.4 What the test steps involve
66 81 832 D3696
D3696.076.01.02.02 05.2019
133
8.4.5
Functional test
1.
Once the above test steps have been completed, restore the
acquisition unit CEREC Primescan AC / Primescan AC / CEREC
Omnicam AC / Omnicam AC to normal operating conditions.
2.
Switch the acquisition unit CEREC Primescan AC / Primescan AC /
CEREC Omnicam AC / Omnicam AC on for the functional test.
3.
Once booting is complete, start the CEREC application.
4.
Take a camera shot of the calibration piece suspended from the
camera.
Ä
The resulting image must not exhibit any surface defects.
The re-test is now complete.
If any of the test criteria have not been met, the acquisition unit CEREC
Primescan AC / Primescan AC / CEREC Omnicam AC / Omnicam AC
must be repaired accordingly.
➢ Record the completion of the test steps in the test report. The test
reports are recorded in a separate document in the relevant national
language and can be accessed via the partner portal by quoting the
material numbers listed below.
Language
Material number
German
6698547
English
6702281
French
6702299
Spanish
6702307
Italian
6702323
Swedish
6702315
Russian
6702331
Danish
6702349
Norwegian
6702356
Dutch
6702364
Finnish
6702372
Japanese
6702380
Portuguese
6702398
Korean
6702406
Chinese
6702414