4 Repair
4.3 Replacing parts (without removing covers)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
96
66 81 832 D3696
D3696.076.01.02.02 05.2019
10. Detach the connector of the power supply from the rear side of the
power supply unit.
Installation
1.
Insert the connector of the power supply into the socket of the rear
side of the power supply unit.
2.
Hang the PC in the track intended for this purpose.
NOTE
Cables can be damaged.
Pay attention to the cables when inserting and moving the PC. The
cables must never become kinked or jammed.
3.
Push the PC completely into the acquisition unit.
4.
Fasten the PC to the acquisition unit with the 3 screws (A).