Calibrating the scanner
1.
Remove the protective cap from the calibration set.
2.
Mount the calibration set on the tip of the scanner until it locks into
place.
3.
Secure the CEREC Primescan / Primescan / CEREC Omnicam /
Omnicam scanner in the calibration set using one hand. Ensure that
the external calibration set screw is fully screwed in a clockwise
motion until it gently locks into place.
4.
Click on the
"OK" button.
Ä
The measuring process starts.
Ä
The software prompts you to proceed to the next latching.
5.
Turn the screw counter-clockwise until you reach the next latching
point.
6.
Click on the
"OK" button. Hold the CEREC Primescan / Primescan /
CEREC Omnicam / Omnicam scanner still.
Ä
The software confirms the calibration process.
Ä
The software prompts you to proceed to the next latching.
7.
Complete steps 5 and 6 a total of 11/17 times.
Ä
The software provides status updates on the calibration and
informs you once the procedure is complete.
Ä
You will be prompted to measure the position of the exit
window.