4 Repair
4.3 Replacing parts (without removing covers)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
66
66 81 832 D3696
D3696.076.01.02.02 05.2019
7.
Release the two screws and remove the strain relief cover (F) from
the camera housing.
IMPORTANT
Ensure that no dust particles penetrate into the interior of the camera.
8.
Carefully disconnect the USB connector (G) from the camera and
remove the old camera cable.
9.
Detach the camera cover (B) from the old camera cable and push it
over the new camera cable.
The old camera cable is no longer required and can be disposed of.
10. Insert the USB connector (G) of the new camera cable into the USB
socket of the camera.
Note: The USB connector in the USB-C Format can be inserted in
both possible orientations.
11. Place the hex (H) of the new camera cable with a flat side
downwards onto the EMC adhesive pad (I) of the camera housing.