6 Maintenance
6.2 Replacing/cleaning the filter
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
128
66 81 832 D3696
D3696.076.01.02.02 05.2019
5.
Open up the lengthways cover by pushing the lateral spring plate.
6.
Remove the metal grille and clean it (e.g. with a vacuum cleaner).
7.
Insert the cleaned metal grille and close the cover shut until the
lateral spring plate latches in.