Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
4 Repair
4.2 Removing covers and holders
66 81 832 D3696
D3696.076.01.02.02 05.2019
59
4.2.3
Replacing the CEREC Omnicam / Omnicam scanner
holder
A
CEREC Omnicam / Omni-
cam scanner holder cover
parts set
B
CEREC Omnicam / Omnicam
scanner holder
C
Heater
D
CEREC Omnicam / Omnicam
internal scanner cable
ü
The service cover is open (see "Opening the service cover [→ 52]").
ü
The right cover is removed (see "Right cover [→ 54]").
1.
Unscrew the scanner from the scanner cable and set the scanner
aside safely.
2.
Unscrew the 3 screws (A) from the device and remove the scanner
holder laterally from the device. When doing so, detach the
connector of the heater cable.
3.
Insert the new scanner holder. When doing this, reinsert the
connector of the heater cable and firmly screw the three screws
onto the device. Ensure that the heater cable does not get crushed.