4 Repair
4.2 Removing covers and holders
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
52
66 81 832 D3696
D3696.076.01.02.02 05.2019
4.2.1.2
Opening the service cover
1.
Turn the turning bolt with a screwdriver (> 5) or another suitable
implement (a coin or suchlike) approx. 90° counterclockwise up to
the stop. The slot of the turning bolt must be in a vertical position in
an opened state.
2.
Press down the service cover as far as it will go and remove it.
4.2.1.3
Closing the service cover
1.
Put the service cover back in position and push it up as far as it will
go.
2.
Turn the turning bolt with a screwdriver (> 5) or another suitable
implement (a coin or suchlike) approx. 90° clockwise up to the stop.
The slot of the turning bolt must be in a horizontal position in a
closed state.
4.2.1.4
Left cover
1.
Unscrew the 3 screws (A, TX 20).