![Dentsply Sirona CEREC Omnicam AC Service Manual Download Page 127](http://html.mh-extra.com/html/dentsply-sirona/cerec-omnicam-ac/cerec-omnicam-ac_service-manual_2483632127.webp)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
6 Maintenance
6.1 Clean the PC fans
66 81 832 D3696
D3696.076.01.02.02 05.2019
127
6
Maintenance
6.1
Clean the PC fans
ü
The PC is removed (see "Replacing the PC including the power
supply unit [→ 94]").
➢ Clean the PC fans using suction from the inside.
6.2
Replacing/cleaning the filter
NOTE
Maintenance performed by trained technical personnel is
recommended on at least an annual basis. Maintenance should
include the filter mat, the filter made from metal grating mesh and the
storage battery.
1.
Flip the footswitch down.
2.
Pull out the filter.
3.
Place a new filter on the footswitch surface.
4.
Close the footswitch again.