5 Settings and adjustments
5.1 Set the monitor kinematics
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
124
66 81 832 D3696
D3696.076.01.02.02 05.2019
5.
Insert the cables into the guides (K) (thin cable first). Ensure that
the radius is adequate (L).
6.
Pivot the cable reliefs above the cables and screw these in with the
2 PT screws (E, F 3 X 8) finger tight. NOTE! Do not tighten too
hard as otherwise the screw bosses may burst!
7.
Carefully retighten the cables. These must run above each other
under the cable guard and may only be pressed slightly by the
cover.
8.
Insert the lever arm cover at the front.