![Dentsply Sirona CEREC Omnicam AC Service Manual Download Page 118](http://html.mh-extra.com/html/dentsply-sirona/cerec-omnicam-ac/cerec-omnicam-ac_service-manual_2483632118.webp)
5 Settings and adjustments
5.1 Set the monitor kinematics
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
118
66 81 832 D3696
D3696.076.01.02.02 05.2019
7.
Loosen 2 PT screws (E, F 3 X 8) and pivot the cable reliefs away.
8.
Remember the cable route (take a photo or mark out the route) so
that you can lay the cable in the exact same way later.
9.
Remove the lever arm cover at the rear.