Important Note (recap):
Always be sure to place the transducer in the same
exact location if it will be removed in between reference lengths and elongation
measurements. This will eliminate any potential sound path error caused by
moving the transducer to a completely different location on the bolt, thus
causing potentially erroneous measurements. Be consistent and as methodical
in your methods as possible. This will help to avoid transducer placement
errors.
1) Press the
UP, DOWN, LEFT, & RIGHT
arrow keys using the left button
wheel to scroll the target cell cursor to the desired storage location.
Note:
Elongation values must be stored in column B - ZZ.
Note:
Do not press the
AUTO SET
key, while measuring elongation’s, as this
activates a high speed mode used specifically with our shut-off box. The
AUTO
SET
feature is only used when measuring reference lengths. If the
AUTO SET
key is accidentally pressed, press it again to turn the high speed mode off.
2) Press the
ENTER
key to save the current elongation in the highlighted cell
location.
Note:
If the measurement was saved in error, press the
CLR
key at anytime to
clear the stored reading and re-measure using steps
1 & 2
above.
3) Continue to repeat this process until all the elongation measurements have
been completed.
Note:
The
Group View Box
can be aborted at any time by pressing the
MEAS
key.