Dakota Ultrasonics
164
Thin, HV Wide, TB 10MHz, TB 5MHz, TB 2MHz, and TB 1MHz. Depending on the
setting selected, the pulser voltage options will include the applicable settings/options
(i.e. if Thin is selected, the max output voltage option is 200V. However if HV Thin is
selected, the max output voltage option is 400V – HV or high voltage, etc.). Note:
the Pulser Voltage menu option is used in conjunction with the Pulse menu option.
Refer to the Pulse menu options to select the desired voltages intended. The
procedure to change the pulser voltage is outlined below:
For quick access and editing, refer to navigating the Hot Menus page 36.
Changing the Pulser Voltage
1) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right and the
ESC
key multiple times to tab left until
the
TUNE
menu is highlighted and displaying the submenu items.
Alternatively, press the
LEFT
and
RIGHT
arrow keys multiple times
using the left button wheel.
2) Use the
UP
and
DOWN
arrow keys multiple times using the left button
wheel to scroll through the sub menu items until
PULSER VOLTAGE
is
highlighted.
3) Use the
LEFT
and
RIGHT
arrow keys multiple times on the right button
wheel to scroll the
Pulser Voltage
options.
4) Once the desired
Pulser Voltage
is displayed, press the
MEAS
key to
return to the measurement screen.