Dakota Ultrasonics
64
Now that a bolt group has been created, and material type selected to save and
document the probe zero data, this next section will go through the process of
obtaining the information needed to use the standard bolt for a probe zero calibration
bar.
Setting up a Bolt Standard – For the First Time
1) With the menu items already active, press the
MENU
key multiple times to
tab right and the
ESC
key multiple times to tab left until the
CAL
menu is
highlighted and displaying the submenu items. Alternatively, press the
LEFT
and
RIGHT
arrow keys multiple times using the left button wheel to
scroll the tabbed menus.
2) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the sub menu items until
ZERO MODE
is
highlighted.
3) Press the
LEFT
and
RIGHT
arrow keys multiple times using the right
button wheel to scroll the zero mode options until the
FIXED
option is
displayed.
Note:
Each time the arrow keys are pressed a confirmation
screen will appear to warn the user that the zero mode is being changed.
Press the
OK
key to advance to the next mode option.
Note:
The fixed option sets the delay factor to zero. In other words, disables
the zero feature altogether.