Dakota Ultrasonics
78
Note:
It is assumed that the temperature sensor is connected to the
MAX II
,
and semi automatic or automatic mode has been selected. The
temperature sensor should also be attached to the specific calibration bar
being measured. If an external pyrometer or other method will be used, the
temperature mode should be set to manual, and the temperature manually
entered into the gauge.
1) With the menu items still active from the previous section, press the
MENU
key multiple times to tab right and the
ESC
key multiple times to tab left until
the
CAL
menu is highlighted and displaying the submenu items.
Alternatively, press the
LEFT
and
RIGHT
arrow keys multiple times
using the left button wheel.
2) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the sub menu items until
ZERO MODE
is
highlighted.
3) Press the
LEFT
and
RIGHT
arrow keys multiple times using the right
button wheel to scroll the zero mode options until the
TWO POINT
option is
displayed. Each time the arrow keys are press, and confirmation screen will
appear letting the user know that the
ZERO MODE
is being changed.
Pressing the
OK
key each time will allow the mode to be changed.
Note:
The two-point option allows the user to use two separate points for the
probe zero calibration.