Dakota Ultrasonics
86
15) Press the
key to automatically setup and measure the 3” cal bar.
Note:
Once an echo is found and the
MAX II
is measuring the cal bar, the
GAIN
, as well as the
POLARITY
may need to be adjusted. Using a large
diameter transducer can produce an extremely strong signal and often the
gain can be turned down to allow the user to see the peaks of the
waveform. A thorough understanding of the section on
“Interpreting the
Waveform”
is very important prior to making any adjustments.
16) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll the hot menu items until
ACCEPT CAL
is highlighted.
17) Press the
ENTER
key to accept the calibration and return to the
Two Point
Zero
list, or
ESC
to cancel the calibration.
18) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the Two Point List items until
ULTRASONIC 2
is highlighted.
19) Apply a drop of couplant to the bolt or transducer and attach it to one end of
the
6” calibration bar
. Use the 3” for the glass block.
20) Press the
ENTER
key to display the measurement screen.
Note:
Rotate the transducer clockwise and counter clockwise applying a small
amount of pressure to eliminate any excess couplant between the
transducer and cal bar surface.