MAX II
Bolt Tension Monitor
53
8) Press the
OK
key to save the
NUM BOLTS
and return to the
Group List
Items
menu, or
ESC
to cancel entering the
NUM BOLTS
.
Selecting the Number of Readings Per Bolt
Note:
A bolt can have up to 51 possible measurements and 1 initial length (L-
REF). There must be at least one reading per bolt. Since this group is being
created to store Zero/Cal Data and verify your probe zero periodically, or before
starting a new application, consider selecting all 51 measurements locations.
1) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the new
Group List Items
until
NUM READS
is highlighted.
2) Press the
ENTER
key to display the
Digits Edit Box
.
3) Press the
UP
and
DOWN
arrow keys multiple times using the left or
right button wheel to scroll the highlighted value.
4) Press the
LEFT
and
RIGHT
arrow keys multiple times using the left or
right button wheel to scroll the digit locations.
5) Repeat steps
3 & 4
until the
NUM READS
value is correctly displayed.
6) Press the
OK
key to save the
NUM READS
and return to the
Group List
Items
menu, or
ESC
to cancel entering the
NUM READS
.