Dakota Ultrasonics
76
as the cursor is moved to a different cell, the display will be updated with the
display view saved with the reading.
4) Scroll the cursor to the next blank cell for the probe zero bolt. This will
always be a cell located in columns
B – ZZ
. As soon as the user has
scrolled to blank cell, the
MAX II
will switch into elongation mode and use all
the data from the probe zero bolt initially stored.
5) Press the
MEAS
key once to activate Hot Menu items. Press the
MEAS
key multiple times to move down and the
ESC
key multiple times to move up
until the
TEMP
cell is highlighted. Alternatively, press the
LEFT
and
RIGHT
arrow keys multiple times using the left button wheel.
6) Press the
LEFT
and
RIGHT
arrow keys multiple times using the right
button wheel to scroll the
TEMP
value.
7) Continue to scroll the temp value until the elongation reads 0.0000. When
the value is at 0.0000 the
MAX II
is officially zeroed and ready to store or
document the zero data prior to making measurements.
8) Press the
ENTER
key to store the 0.0000 elongation value. Why bother
storing the 0.0000 elongation value? Because the adjustment to
temperature is also stored with the measurement. This documents the
changes for reporting requirements.