11) Measure and record the Ultrasonic Length at Load 2 by pressing the
ENTER
key. The
Digits Edit Box
will be displayed.
Note:
Load 2 should be stored in column C.
12) Enter the
Known Load
value by pressing the
UP
and
DOWN
arrow
keys multiple times using the left or right button wheel to scroll the
highlighted value.
13) Press the
LEFT
and
RIGHT
arrow keys multiple times using the left or
right button wheel to scroll the digit locations.
14) Repeat steps
8 & 9
until the
Known Load
value is correctly displayed.
15) Press the
OK
key to store the
Known Load
and return to the measurement
screen, or
ESC
to cancel entering the
Known Load
.
16)
Increase the applied load to approximately the maximum load, which is to
be placed on the bolt under actual working conditions.
17) Measure and record the Ultrasonic Length at Load 3 by pressing the
ENTER
key. The
Digits Edit Box
will be displayed.
Note:
Load 3 should be stored in column D.
18) Enter the
Known Load
value by pressing the
UP
and
DOWN
arrow
keys multiple times using the left or right button wheel to scroll the
highlighted value.
19) Press the
LEFT
and
RIGHT
arrow keys multiple times using the left
button wheel to scroll the digit locations.
20) Repeat steps
18 & 19
until the
Known Load
value is correctly displayed.
21) Press the
OK
key to store the
Known Load
and return to the measurement
screen, or
ESC
to cancel entering the
Known Load
.
Summary of Contents for MAX II
Page 2: ......
Page 7: ......
Page 8: ......
Page 54: ...Dakota Ultrasonics 46 Figure 6 Reflection in a bending bolt ...