Dakota Ultrasonics
82
Note:
The default is
AUTO
, allowing the
MAX II
the ability to select either
positive or negative polarity. However, the polarity can be set and fixed to a
specific polarity, positive/negative, as selected by the user.
31) Press the
LEFT
and
RIGHT
arrow keys multiple times using the right
button wheel to scroll
AUTO
,
POSITIVE
or
NEGATIVE
polarity options.
32) Press the
ESC
key once to leave the
Two Point Zero
list.
33) Press the
MENU
key multiple times to tab right and the
ESC
key multiple
times to tab left until the
MATL
menu is highlighted and displaying the
submenu items. Alternatively, press the
LEFT
and
RIGHT
arrow keys
multiple times using the left button wheel.
Note:
In order for the
MAX II
to display the correct measurement value during
calibration, the material velocity and temperature coefficient must be set to the
same values as in the calibration section, and the stress factor set to 1.000.
34) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the sub menu items until
VELOCITY
is
highlighted.
35) Press the
ENTER
key to display the
Digits Edit Box
.
36) Press the
UP
and
DOWN
arrow keys multiple times using the left or
right button wheel to scroll the highlighted value.
Summary of Contents for MAX II