MAX II
Bolt Tension Monitor
81
20) Press the
UP
and
DOWN
arrow keys multiple times using the left or
right button wheel to scroll the highlighted value.
21) Press the
LEFT
and
RIGHT
arrow keys multiple times using the left or
right button wheel to scroll the digit locations.
22) Repeat steps
20 & 21
until the
VELOCITY
value is correctly displayed.
23) Press the
OK
key to return to the
Two Point Zero
list, or
ESC
to cancel
entering the
VELOCITY
.
24) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the
Two Point Zero
list until
TEMP COEF
is
highlighted.
Note:
This should be set to 55 (99 metric) for the 3 & 6” bars supplied by
Dakota Ultrasonics. If the glass block is used, it should be set to 0. The reason
for this is because glass has a very low temperature coefficient.
25) Press the
ENTER
key to display the
Digits Edit Box
.
26) Press the
UP
and
DOWN
arrow keys multiple times using the left or
right button wheel to scroll the highlighted value.
27) Press the
LEFT
and
RIGHT
arrow keys multiple times using the left or
right button wheel to scroll the digit locations.
28) Repeat steps
26 & 27
until the
TEMP COEF
value is correctly displayed.
29) Press the
OK
key to return to the
Two Point Zero
list, or
ESC
to cancel
entering the
TEMP COEF
.
30) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the
Two Point Zero
list until
POLARITY
is
highlighted.