MAX II
Bolt Tension Monitor
79
Note:
The one point option can optionally be selected if only one cal bar will be
used. Although the procedures below are described with the two-point option in
mind, they will also work in the case of a one-point calibration.
4) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the sub menu items until
MEASURE ZERO
is
highlighted.
5) Press the
ENTER
key to display the
Two Point Zero
list box.
6) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the
Two Point Zero
list until
PHY LEN 1
is
highlighted.
Note:
PHY LEN 1
is the actual physical length of the first calibration standard
noted on the certificate included with the bars. If the glass block is being used,
the 1” side should be used for
PHY LEN 1
.
7) Press the
ENTER
key to display the
Digits Edit Box
.
8) Press the
UP
and
DOWN
arrow keys multiple times using the left or
right button wheel to scroll the highlighted value.
9) Press the
LEFT
and
RIGHT
arrow keys multiple times using the left or
right button wheel to scroll the digit locations.
10) Repeat steps
8 & 9
until the
PHY LEN 1
value is correctly displayed.