MAX II
Bolt Tension Monitor
85
5) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the sub menu items until
OPEN
is highlighted.
6) Press the
ENTER
key to display the list of log files.
7) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the log names until
CAL DATA
is highlighted.
8) Press the
ENTER
key to open the log, followed by pressing the
OK
to
confirm “load from memory”, or
ESC
to cancel opening the log file.
9) Press the
MENU
key multiple times to tab right and the
ESC
key multiple
times to tab left until the
CAL
menu is highlighted and displaying the sub
menu items. Alternatively, press the
LEFT
and
RIGHT
arrow keys
multiple times using the left button wheel.
10) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the sub menu items until
MEASURE ZERO
is
highlighted.
11) Press the
ENTER
key to display the
Two Point Zero
list.
12) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the sub menu items until
ULTRASONIC 1
is
highlighted.
13) Apply a drop of couplant to the standard or transducer and attach it to one
end of the
3” calibration bar
.
14) Press the
ENTER
key to display the measurement screen.
Note:
Rotate the transducer clockwise and counter clockwise applying a small
amount of pressure to eliminate any excess couplant between the
transducer and cal bar surface.