Dakota Ultrasonics
80
11) Press the
OK
key to return to the
Two Point Zero
list, or
ESC
to cancel
entering the
PHY LEN 1
.
12) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the
Two Point Zero
list until
PHY LEN 2
is
highlighted.
Note:
PHY LEN 2
is the actual physical length of the second calibration
standard. If the glass block is being used, the 3” side should be used for
PHY
LEN 2
.
13) Press the
ENTER
key to display the
Digits Edit Box
.
14) Press the
UP
and
DOWN
arrow keys multiple times using the left or
right button wheel to scroll the highlighted value.
15) Press the
LEFT
and
RIGHT
arrow keys multiple times using the left or
right button wheel to scroll the digit locations.
16) Repeat steps
14 & 15
until the
PHY LEN 2
value is correctly displayed.
17) Press the
OK
key to return to the
Two Point Zero
list, or
ESC
to cancel
entering the
PHY LEN 2
.
18) Press the
UP
and
DOWN
arrow keys to scroll through the
Two Point Zero
list until
VELOCITY
is highlighted.
Note:
This refers the velocity of the probe zero cal bars. Since there is a
velocity for each cal bar, use the velocity of the longest cal bar as it will be most
significant – 6” bar. If the glass block is being used, it will have only one
velocity specified for the block.
19) Press the
ENTER
key to display the
Digits Edit Box
.