Dakota Ultrasonics
58
2) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the sub menu items until
ZERO MODE
is
highlighted.
3) Press the
LEFT
and
RIGHT
arrow keys multiple times using the right
button wheel to scroll the zero mode options until the
AUTO
option is
displayed.
Note:
The auto option will perform a two-point calibration option allows the
user to use two separate points for the probe zero calibration.
4) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the sub menu items until
MEASURE ZERO
is
highlighted.
5) Press the
ENTER
key to display the
One Point Zero
list box.
6) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the
One Point Zero
list until
PHY LENGTH
is
highlighted.
Note:
PHY LENGTH
is the actual physical length of the calibration standard.
7) Press the
ENTER
key to display the
Digits Edit Box
.
8) Press the
UP
and
DOWN
arrow keys multiple times using the left or
right button wheel to scroll the highlighted value.