Dakota Ultrasonics
122
1) Press the
MENU
key multiple times to tab right and the
ESC
key multiple
times to tab left until the
CAL
menu is highlighted and displaying the
submenu items. Alternatively, press the
LEFT
and
RIGHT
arrow keys
multiple times using the left button wheel.
2) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the sub menu items until
LOAD CAL MODE
is highlighted.
3) Press the
LEFT
and
RIGHT
arrow keys multiple times using the right
button wheel to scroll through the unit options
OFF
,
VECTOR
(zero offset),
REGRESSION
, until the correct mode is displayed.
4) Press the
MEAS
key to return to the measurement mode.
Now that a group has been created and activated, and the load calibration mode
selected, we’re ready to perform the field calibration.
Performing a Field Calibration
1) Place the first sample bolt in the load apparatus.
Note:
The following steps assume that the user has previously selected a
similar material type for the list of materials and the approximate length of the
bolt has been entered. Therefore, the
MAX II
is currently setup and ready to
measure in elongation mode.